DC Volts and DC Current Gain Verification

This procedure verifies DC volts and DC current gain.
1. Make sure you have read
2. Connect the calibrator to the input terminals. For the 34461A/65A/70A, use the front panel input ter-
minals and select the front input terminals with the Front/Rear switch.
3. Configure each function and range in the order shown in the table below corresponding to the DMM
model number. Provide the input shown in the table.
4. Make a measurement and return the result. Compare measurement results to the test limits shown
below. (Be certain to allow for appropriate source settling.)
Input
Voltage Function
100 mV
-100 mV
1 V
-1 V
4 V
10 V
DC Volts
-10 V
100 V
-100 V
1000 V
-500 V
[1] Select 10 PLC.
Q
: Quick performance verification test point.
Input
Current Function
2 A
1 A
100 mA
DC Current
10 mA
1 mA
100 µA
[1] Select 10 PLC.
Q
: Quick performance verification test point.
510
Test
Considerations.
Values for 34460A DMM
Quick
[1]
Range
24 hours 90 days
Check
100 mV
± 10 µV
1 V
± 39 µV
10 V
± 140 µV
Q
10 V
± 290 µV
100 V
± 3.6 mV
Q
± 36 mV
1000 V
± 21 mV ± 42.5 mV ± 52.5 mV ± 62.5 mV
Values for 34460A DMM
Quick
[1]
Range
24 hours
Check
3 A
± 4.2 mA
1 A
Q
± 560 µA
100 mA
± 14 µA
10 mA
Q
± 2.7 µA
1 mA
± 0.13 µA ± 0.36 µA
100 µA
± 0.03 µA ± 0.065 µA ± 0.075 µA ± 0.085 µA
Error from Nominal
1 year
± 13.5 µV ± 15.5 µV
± 70 µV
± 90 µV
± 250 µV
± 350 µV
± 550 µV
± 800 µV
± 7.1 mV
± 9.1 mV ± 11.6 mV
± 75 mV
± 95 mV
± 120 mV
Error from Nominal
90 days
1 year
± 4.6 mA
± 4.6 mA
± 900 µA
± 1.1 mA
± 35 µA
± 55 µA
± 5 µA
± 7 µA
± 0.56 µA
Keysight True volt Series Operating and Service Guide
2 years
± 18 µV
± 115 µV
± 450 µV
± 1.1 mV
2 years
± 5.2 mA
± 1.3 mA
± 65 µA
± 8 µA
± 0.66 µA