Keysight Truevolt Series Operating And Service Manual page 213

Digital multimeters
Table of Contents

Advertisement

SCPI Programming Reference
CALCulate:LIMit:CLEar[:IMMediate]
Clears front panel indications of limits being exceeded and clears bit 11 ("Lower Limit Failed") and bit 12
("Upper Limit Failed") in the Condition Register of the Questionable Data Register event register group.
The corresponding event register bits are unaffected.
A condition register continuously monitors the state of the instrument. Condition register bits are updated
in real time; they are neither latched nor buffered.
An event register is a read-only register that latches events from the condition register. While an event bit
is set, subsequent events corresponding to that bit are ignored.
Parameter Typical Return
(none)
(none)
Clear the limit test results:
CALC:LIM:CLE
Does not clear measurements in reading memory.
l
The instrument clears front panel indications of limits being exceeded and clears bits 11 and 12 in the
l
Questionable Data Register when the measurement function changes, or when any of the following
commands is executed:
CALCulate:LIMit:STATe ON
INITiate
MEASure:< function >?
READ?
CALCulate:LIMit:CLEar
*RST
SYSTem:PRESet
To clear statistics, limits, histogram data, and measurement data, use CALCulate:CLEar[:IMMediate].
l
214
Keysight True volt Series Operating and Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

34461a34465a34470a34460a