SCPI Programming Reference
SAMPle Subsystem
The SAMPle commands are used with the TRIGger commands to set up the triggering parameters. Use the
SAMPle commands to set a sample count, sample source, and delay time for triggered sampling.
It is recommended that all triggered measurements be made using an appropriate fixed
manual range. That is, turn autorange off
a fixed range using the SENSe:<function>:RANGe, CONFigure, or
Command Summary
SAMPle:COUNt
SAMPle:COUNt:PRETrigger
SAMPle:SOURce
SAMPle:TIMer
SAMPle:COUNt {< count >|MIN|MAX|DEF}
SAMPle:COUNt? [{MIN|MAX|DEF}]
Specifies the number of measurements (samples) the instrument takes per trigger.
6
1 to 1,000,000 (1x10
). Default: 1. (34460A/61A)
1 to 1,000,000,000 (1x10
Return ten sets of four resistance measurements, using an external trigger to
start each measurement set:
CONF:RES 1E6
SAMP:COUN 4
TRIG:COUN 10
TRIG:SOUR EXT;SLOP NEG
READ?
Typical Response: +1.00520000E+06, ... (40 measurements)
The front panel sample annunciator ( " * " ) turns on during each measurement at slower rates. The
l
annunciator toggles at a fixed rate for fast measurements.
You can use the specified sample count in conjunction with a trigger count (TRIGger:COUNt), which
l
sets the number of triggers to be accepted before returning to the "idle" trigger state. The total num-
ber of measurements returned are the product of the sample count and trigger count.
You can store up to 1,000 measurements in the reading memory of the 34460A, 10,000 meas-
l
urements on the 34461A, 50,000 measurements on the 34465A/70A ( without the MEM option), or
2,000,000 measurements on the 34465A/70A ( with the MEM option). If reading memory overflows,
new measurements overwrite the oldest measurements stored; the most recent measurements are
always preserved. No error is generated, but the Reading Mem Ovfl bit (bit 14) is set in the Ques-
tionable Data Register's condition register (see
306
Parameter
9
). Default: 1. (34465A/70A)
Status System
(SENSe:<function>:RANGe:AUTO
MEASure
Typical
Return
+1
Introduction).
Keysight True volt Series Operating and Service Guide
OFF), or set
command.