High Current Verification

This procedure applies to the 34461A, 34465A, and 34470A only (not applicable to the 34460A).
This procedure verifies AC and DC current on the 10 A terminals.
1. Make sure you have read
2. Connect the calibrator to the front 10 A input terminals and select the Front input terminals with the
Front/Rear switch.
3. Select each function in the order shown below. Provide the indicated input current and frequency, and
for ACI, use the 200 Hz input filter. Compare measurement results to the test limits shown below. (Be
certain to allow for appropriate source settling.)
Current
Function
[1]
DCI
5 A
DCI
10 A
ACI
10 A
[1]
If you are NOT using the Fluke 5725 for calibration, reduce the current to 2 A.
Keysight True volt Series Operating and Service Guide
Test
Considerations.
Values for the 34461A, 34465A, and 34470A DMMs
Input
Frequency Range
N/A
10 A
N/A
10 A
5 kHz
10 A
Error from Nominal
Quick
24 hours
Check
± 3.5 mA
Q
± 16 mA
± 24 mA (34465A/70A)
± 29 mA (34461A)
Performance Verification
90
1
2
days
year
years
± 8.5
± 7 mA
mA
± 26
± 23 mA
mA
± 29 mA