The following table lists the bit definitions for the Questionable Data Register :
The overload bits are set once per
new INITiate is sent.
Bit Num-
Bit Name
Decimal
ber
0
Voltage Overload
1
Current Overload
2
Sample Timing
Violation
3
Not Used
4
Temperature Over-
load
5
Frequency Over-
load/Underflow
6
Not Used
7
Not Used
8
Calibration Corrupt
9
Resistance Over-
load
10
Capacitance
Overload
11
Lower Limit Failed
12
Upper Limit Failed
13
Not Used
14
Memory Overflow
15
Not Used
Keysight True volt Series Operating and Service Guide
INITiate
Value
1
Only reported as event. Returns 0. Read the Event Register.
Only reported as event. Returns 0. Read the Event Register.
2
4
A sample timing violation has occurred involving the
mand. Sample timing may not be valid. Applies only to the 34465A and
34470A.
8
(Reserved for future use)
16
Only reported as event. Returns 0. Read the Event Register.
32
Only reported as event. Returns 0. Read the Event Register.
64
(Reserved for future use)
128
(Reserved for future use)
256
At least one calibration constant is corrupt.
512
Only reported as event. Returns 0. Read the Event Register.
1024
Only reported as event. Returns 0. Read the Event Register.
2048
The most recent measurement failed the lower limit test.
4096
The most recent measurement failed the upper limit test.
8192
(Reserved for future use)
16384
Reading memory is full. One or more (oldest) measurements have been lost.
32768
(Reserved for future use)
command. If you clear an overload bit, it is not set again until a
Definition
SCPI Programming Reference
com-
SAMPle:TIMer
387