SCPI Programming Reference
CONFigure[:VOLTage]:{AC|DC} [{< range >|AUTO|MIN|MAX|DEF} [, {< resolution >|MIN|MAX|DEF}]]
Sets all measurement parameters and trigger parameters to their
urements. Also specifies the range and resolution.
The maximum range parameter (MAX) is 1000 V. However, the SAFETY LIMIT on the front
and rear HI/LO input terminals is 750 VAC (rms). The rms voltage is waveform dependent.
A sine wave is limited to 750 VAC (rms), but a 1000 Vpk square wave is safe. Connection
to AC mains power is further limited to CAT II (300 V). See
Information
instrument.
<range>: {100 mV|1 V|10 V|100 V|1000 V}. Default: AUTO (autorange).
<resolution> (AC): optional and ignored; fixed at 6½ digits.
Resolution Table
<resolution> (DC): See
default is equivalent to 10 PLC.
Specify <resolution> in the measurement's units (V, A, Hz, Ω, and so on).
Configure AC voltage measurements using the 100 V range. Make and read two measurements:
CONF:VOLT:AC 100
SAMP:COUN 2
READ?
Typical Response: +8.54530000E+01,+8.54520000E+01
Configure DC voltage measurements using the 10 V range with 1 mV resolution. Then make a meas-
urement using an external trigger with positive slope (rising edge) and read the measurement:
CONF:VOLT:DC 10,0.001
TRIG:SOUR EXT;SLOP POS
INIT
FETC?
Typical Response: +4.27150000E+00
You can let autoranging select the measurement range, or you can manually select a fixed range.
l
Autoranging conveniently selects the range for each measurement based on the input signal. For fast-
est measurements, use manual ranging (autoranging may require additional time for range selection).
Autoranging (AUTO or DEFault), will generate an error if you specify a < resolution > because the instru-
l
ment cannot accurately resolve the integration time (especially if the input continuously changes). If
your application requires autoranging, specify DEFault for the < resolution > or omit the < resolution >
altogether.
Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of
l
range.
If the input signal is greater than can be measured on the specified manual range, the instrument dis-
l
plays the word Overload on front panel and returns "9.9E37" from the remote interface.
252
for more information on the safety features and the safe operation of this
Parameter
Range, Resolution and NPLC.
or
default values
for AC or DC voltage meas-
Safety and Regulatory
The
Keysight True volt Series Operating and Service Guide
Typical Return
(none)