Keysight Truevolt Series Operating And Service Manual page 256

Digital multimeters
Table of Contents

Advertisement

DATA:POINts:EVENt:THReshold < count >
DATA:POINts:EVENt:THReshold?
Sets the number of measurements to store in reading memory before setting bit 9 in the Standard Oper-
ation Register group event register to 1. An event register is a read-only register that latches events from
the condition register. While an event bit is set, subsequent events corresponding to that bit are ignored.
For more information on the SCPI Status System, see
34460A: 1 to 1,000 readings
34461A: 1 to 10,000 readings
34465A/70A: 1 to 50,000 readings (without the MEM option), or 1 to 2,000,000
readings (with the MEM option) 
The default value is 1 for both models.
Set the reading memory threshold to 125 measurements:
DATA:POIN:EVEN:THR 125
To report any subsequent events, the measurement count must first drop below the programmed
l
threshold before reaching the threshold again.
Use
STATus:OPERation:ENABle
l
Event register) to be reported to the Status Byte.
Once the Memory Threshold bit (bit 9 in the Standard Operation Event register) is set, it remains set
l
until cleared by
STATus:OPERation:EVENt?
The Standard Operation Condition register shows the current condition. See
l
duction
for further information.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
Keysight True volt Series Operating and Service Guide
Status Subsystem
Parameter
to enable the Memory Threshold bit (bit 9 in the Standard Operation
or *CLS.
SCPI Programming Reference
Introduction.
Typical Return
+125
STATus Subsystem Intro-
257