Keysight Truevolt Series Operating And Service Manual page 506

Digital multimeters
Table of Contents

Advertisement

Input Function
Open
DC Current
Short
DC Volts
Short 4-wire Ohms
[1] Select 10 PLC.
Q
: Quick performance verification test point.
Keysight True volt Series Operating and Service Guide
Values for 34461A DMM
Quick
[1]
Range
24 hours 90 days 1 year 2 years
Check
100 µA
Q
± 0.02 µA
1 mA
10 mA
100 mA
± 4 µA
1 A
± 60 µA
3 A
10 A
100 mV
Q
± 3 µV
1 V
± 6 µV
10 V
± 40 µV
100 V
1000 V
± 6 mV
100 Ω
± 3 mΩ
1 kΩ
± 5 mΩ
10 kΩ
Q
± 50 mΩ
100 kΩ
± 500 mΩ
1 MΩ
10 MΩ
Error from Nominal
± 0.025 µA
± 0.060 µA
± 2 µA
± 5 µA
± 100 µA
± 600 µA
± 1 mA
± 3.5 µV
± 7 µV
± 50 µV
± 600 µV
± 10 mV
± 4 mΩ
± 10 mΩ
± 100 mΩ
± 1 Ω
± 10 Ω
± 100 Ω
Performance Verification