Keysight Truevolt Series Operating And Service Manual page 317

Digital multimeters
Table of Contents

Advertisement

SCPI Programming Reference
[SENSe:]CAPacitance:SECondary {"OFF"|"CALCulate:DATA"}
[SENSe:]CAPacitance:SECondary?
Selects the secondary measurement function for capacitance measurements.
Parameter
{"OFF"|"CALCulate:DATA"}, default "OFF"
Configure capacitance and secondary measurements. Make capacitance measurement and retrieve sec-
ondary measurement.
CONF:CAP AUTO,MAX
CAP:SEC "CALC:DATA"
READ?;DATA2?
Typical Response: +3.01566373E-10;+3.01566373E-10
"CALCulate:DATA" - The measurement value before any math operations are done (including NULL).
l
Use
READ?
or
INITiate
l
urement.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
318
to start the measurement. Use
Typical Return
"CALC:DATA"
[SENSe:]DATA2?
to retrieve the secondary meas-
Keysight True volt Series Operating and Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

34461a34465a34470a34460a

Table of Contents