Keysight Truevolt Series Operating And Service Manual page 28

Digital multimeters
Table of Contents

Advertisement

IP Address
You can enter a static IP address for the instrument as a four-byte integer expressed in dot notation. Each
byte is a decimal value, with no leading zeros (for example, 169.254.2.20).
If DHCP is on, it attempts to assign an IP address to the instrument. If it fails, Auto-IP attempts to
l
assign an IP address to the instrument.
Contact your LAN administrator to obtain an IP address.
l
This setting is non-volatile; it will not be changed by power cycling or
l
Front Panel:[Utility] > I/O Config > LAN Settings > Modify Settings
l
Then set the first softkey to Manual and press IP Address to enter a new IP address.
SCPI:
SYSTem:COMMunicate:LAN:IPADdress "< address >"
l
If you change this parameter, you must either press the Apply Changes softkey (front panel) or send
l
SYSTem:COMMunicate:LAN:UPDate
Subnet Mask
Subnetting allows the LAN administrator to subdivide a network to simplify administration and minimize
network traffic. The subnet mask indicates the portion of the host address used to indicate the subnet.
Contact your LAN administrator for details.
l
This setting is non-volatile; it will not be changed by power cycling or
l
Front Panel:[Utility] > I/O Config > LAN Settings > Modify Settings
l
Then set the first softkey to Manual and press Subnet Mask to enter a new subnet mask with the
arrow keys (for example: 255.255.0.0).
SCPI:
SYSTem:COMMunicate:LAN:SMASk "< mask >"
l
If you change this parameter, you must either press the Apply Changes softkey (front panel) or send
l
SYSTem:COMMunicate:LAN:UPDate
Keysight True volt Series Operating and Service Guide
(remote interface) for the change to take effect.
(remote interface) for the change to take effect.
Introduction to the Instrument
*RST
or SYSTem:PRESet.
*RST
or SYSTem:PRESet.