SYSTem:COMMunicate:LAN:HOSTname "< name >"
SYSTem:COMMunicate:LAN:HOSTname? [{CURRent|STATic}]]
Assigns a hostname to the instrument. A hostname is the host portion of the domain name, which is trans-
lated into an IP address. If Dynamic Domain Name System (Dynamic DNS) is available on your network
and your instrument uses DHCP, the hostname is registered with the Dynamic DNS service at power-on. If
DHCP is enabled (SYSTem:COMMunicate:LAN:DHCP ON), the DHCP server can change the specified host-
name.
If you change this setting, you must send
ate the new setting.
String of up to 15 characters.
Must start with letter (A-Z)
May contain letters, numbers (0-9), or dashes ("-")
Default: "K-<instrument model number>-nnnnn", where nnnnn is the last five digits of the
instrument's serial number.
Define a hostname:
SYST:COMM:LAN:HOST "LAB1-DMM"
SYST:COMM:LAN:UPD
If no hostname exists, a null string ("") is returned.
l
On the 34460A, this requires option 34460A-LAN or option 3446LANU.
l
For the query form, specify "CURRent" (default) to read the value currently being used by the instru-
l
ment. Specify "STATic" to read the host name currently stored in nonvolatile memory within the instru-
ment (may not be the actual name used by the instrument if DHCP is enabled).
This setting is non-volatile; it is not changed by power cycling, a Factory Reset (*RST), or an Instru-
l
ment Preset (SYSTem:PRESet).
This parameter is set to its default value when the instrument is shipped from the factory and after
l
SYSTem:SECurity:IMMediate.
Keysight True volt Series Operating and Service Guide
SYSTem:COMMunicate:LAN:UPDate
Parameter
SCPI Programming Reference
to activ-
Typical
Return
"LAB1-
34461A"
411