Calibration Security

This section describes the instrument's calibration security system. If you forget the security code, you can
override calibration
security.
Security Overview
A security code (calibration password) prevents accidental or unauthorized instrument calibration.
When shipped from the factory, the instrument is secured, with the security code (calibration pass-
l
word) set to AT3446XA .
The same security code must be used for front panel and remote operation. If you secure the instru-
l
ment from the front panel, use that same code to unsecure it from the remote interface.
This setting is non-volatile; it will not be changed by power cycling or
l
Security code rules:
l
Unquoted string up to 12 characters
Must start with letter (A-Z)
May contain letters, numbers (0-9) and underscores
Front Panel: [Utility] > Test / Admin > Security
l
SCPI:
CALibration:SECure:STATe
l
Keysight True volt Series Operating and Service Guide
Calibration Adjustment Procedures
*RST
or SYSTem:PRESet.