General Disassembly Procedure - Keysight Truevolt Series Operating And Service Manual

Digital multimeters
Table of Contents

Advertisement

General Disassembly Procedure

1. Turn off the power and remove all measurement leads and other cables, including the power cord,
from the instrument before continuing.
2. Rotate the handle to the upright position and remove it by pulling outward where it attaches to the
case.
3. Pull off the front rubber bumper.
4. Unscrew the two captive screws in the rear bezel and remove the rear bezel and rubber bumper.
5. Remove the screw on the bottom of the instrument and place it in a safe location for re-assembly.
Slide off the instrument cover.
At this point, the instrument it sufficiently disassembled for power supply troubleshoot-
ing purposes. You could also change the battery or
point. If you need to disassemble the front panel, continue with the steps shown below.
Keysight True volt Series Operating and Service Guide
Service and Repair
override calibration security
at this