Dc Voltage - Keysight Truevolt Series Operating And Service Manual

Digital multimeters
Table of Contents

Advertisement

DC Voltage

This section describes how to configure DC voltage measurements from the front panel, including DCV
ratio measurements.
Step 1: Configure the test leads as shown.
Step 2: Press [DCV] on the front panel.
Step 3:
For the 34460A/61A , press Aperture and choose the number of power-line cycles (PLCs) to use for
l
the measurement. Only 1, 10, and 100 PLC provide normal mode (line frequency noise) rejection.
Selecting 100 PLC provides the best noise rejection and resolution, but the slowest measurements:
Keysight True volt Series Operating and Service Guide
Features and Functions
57