Trigger Source; Device Actions - Keysight B2980B Series User Manual

Femto/picoammeter electrometer/high resistance meter
Table of Contents

Advertisement

Trigger Source

B2980 supports the following trigger sources. The trigger source must be set to
the ARM event and the TRIGGER event individually.

Device Actions

B2980 supports the following device actions. Also see
Keysight B2980B User's Guide, Edition 1
AUTO (automatic internal, AINT): Trigger source best suited for the present
operating mode is automatically selected by the internal algorithms.
BUS: Remote interface trigger command such as GET, TRIGger, and *TRG
TIMER (timer, TIMer): Signal internally generated every interval set by the
Period field
INT1 or INT2 (internal): Signal from the internal bus 1 or 2
EXT1, EXT2, EXT3, EXT4, EXT5, EXT6, or EXT7 (external): Signal from the DIO
pin n, which is an I/O port of the Digital I/O D-sub connector on the rear panel.
n = 1 to 7
LAN: LXI trigger
TIN: Signal from the Trigger Input connector
Transient device action (voltage source output for B2985B/B2987B)
Voltage source channel starts voltage output for applying one output value if
the TRIGGER event was detected and the Source delay time elapsed.
Acquire device action (measurement)
Measurement channel starts measurement for obtaining one measurement
data if the TRIGGER event was detected and the Measure delay time elapsed.
Function Details
Trigger System
Figure
5-11.
239

Hide quick links:

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

B2981bB2983bB2985bB2987b

Table of Contents