TRIGger:DELay {< seconds >|MIN|MAX|DEF}
TRIGger:DELay? [{MIN|MAX|DEF}]
Sets the delay between the trigger signal and the first measurement. This may be useful in applications
where you want to allow the input to settle before taking a measurement or for pacing a burst of meas-
urements.
Selecting a specific trigger delay disables the automatic trigger delay
For AC measurements (voltage or current), there is a relationship between the bandwidth
setting
([SENSe:]VOLTage:AC:BANDwidth
[SENSe:]CURRent:AC:BANDwidth) and the default trigger delay setting. This delay
ensures that there is sufficient settling time for the AC measurement. See
Trigger Delays
Parameter
0 to ~3600 seconds (~1 µs steps). Default: 1 s.
Step size for DC measurements is approximately 1 µs.
Step size depends on AC bandwidth for AC measurements.
Return five DC voltage measurements. There is a 2 second delay before each measurement.
CONF:VOLT:DC 10
SAMP:COUN 5
TRIG:DEL 2
READ?
Typical Response:
+4.27230000E+00,+4.27150000E+00,+4.27190000E+00,+4.27170000E+00,+4.27200000E+00
Due to internal quantization, the actual delay that you set may be slightly different than your specified
l
value. The increment is approximately 1 µs.
By default,
TRIGger:DELay:AUTO
l
function, range and integration time (see
delay longer than the automatic delay for long cables, high capacitance or high impedance signals.
If you specify a trigger delay with this command, that delay is used for all functions ( except
l
ity
and DIODe) and ranges. The CONTinuity and DIODe tests ignore the trigger delay setting.
If you have configured the instrument for more than one measurement per trigger
l
>1), the delay is inserted after the trigger and between consecutive measurements.
The instrument selects automatic trigger delay after a Factory Reset (*RST) or an Instrument Preset
l
(SYSTem:PRESet).
Keysight True volt Series Operating and Service Guide
for further information.
is ON. The instrument automatically determines the delay based on
Automatic Trigger
SCPI Programming Reference
(TRIGger:DELay:AUTO
or
Typical Return
+2.00000806E+00
Delays). However, you may need to set a
OFF).
Automatic
CONTinu-
(SAMPle:COUNt
429