To Refill The Ei Calibration Vial - Agilent Technologies 7250 Troubleshooting And Maintenance Manual

Accurate-mass quadrupole time-of-flight gc/ms system
Hide thumbs Also See for 7250:
Table of Contents

Advertisement

4
General Maintenance

To Refill the EI Calibration Vial

Materials needed
Procedure
1 Stop any tuning or data acquisition.
2 Turn off the MS electronics.
3 Remove the source window cover. See the
4 Turn the calibration vial collar counterclockwise to loosen it. (See
5 Pull the calibration vial out. You may feel some resistance due to the O-ring
Figure 7.
7250 Accurate-Mass Q-TOF Troubleshooting and Maintenance Manual
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
PFTBA (05971-60571)
Q-TOF GC/MS System Operating
Do not remove the collar.
around the vial tube section.
Calibration vial
Collar
Removing the EI calibration vial
Agilent 7250 Accurate-Mass
Manual.
Figure
7.)
73