Wavelength Calibration Procedure; Preparation Of The Glycogen Calibration Sample - Agilent Technologies 1200 Series User Manual

Fluorescence detector
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

Wavelength Calibration Procedure

When required
Tools required
Parts required
Steps

1 Preparation of the Glycogen Calibration Sample.

2 Preparation of the Flow Cell.
3 Wavelength Calibration.
Preparation of the Glycogen Calibration Sample
1 To prepare 10 ml of the calibration solution you have to use 10 mg of the
2 Fill the prepared amount into a suitable bottle/vial.
3 Fill 10 ml of distilled water into the vial and shake.
4 Wait 5 minutes and shake again. After 10 minutes the solution is ready.
1200 Series FLD User Manual
If application requires, or after replacement of flow cell or lamp.
Laboratory balance
Glycogen Calibration Sample, Syringe, needle, sample filter and PEEK fitting
from the Accessory Kit, see
Glycogen sample (a tolerance of
"Accessory Kit"
on page 135
±
20% is not critical).