Wavelength calibration setting failed

Error ID: 7310
The intensity maximum was not found during wavelength calibration.
Calibration 0 Failed:
Calibration 1 Failed:
Probable cause
1
2
3
4
5
6
7
Agilent 1260 Infinity VWD User Manual
Lamp is OFF.
Incorrect flow cell installation.
Flow cell contamination or air bubbles.
Intensity too low.
Current step value too far from maximum.
Misaligned/defective grating assembly.
Defective main board.
Error Information
Detector Error Messages
Zero-order calibration failed.
656 nm calibration failed.
Suggested actions
Switch on the lamp.
Ensure the flow cell is installed correctly.
Clean/replace flow cell windows or remove air
bubbles.
Replace lamp.
• Repeat the calibration.
• Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.