Wavelength Recalibration Lost

Error ID: 1037
The calibration information needed for your detector to operate correctly has
been lost.
During calibration of the detector the calibration values are stored in ROM. If
no data is available in the spectrometer ROM, the error message is generated.
Probable cause
1
2

Illegal Temperature Value from Sensor on Main Board

Error ID: 1071
This temperature sensor (located on the detector main board) delivered a
value outside the allowed range. The parameter of this event equals the
measured temperature in 1/100 centigrade. As a result the temperature
control is switched off.
Probable cause
1
2
Agilent 1200 Infinity Series DAD User Manual
The detector is new.
The detector has been repaired.
Defective sensor or main board.
Detector is exposed to illegal ambient
conditions.
Error Information
Detector Error Messages
Suggested actions
Recalibrate the detector.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Verify that the ambient conditions are within
the allowed range.