Wavelength Calibration Setting Failed - Agilent Technologies InfinityLab LC Series User Manual

Variable wavelength detectors
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

7
Error Information
Detector Error Messages

Wavelength calibration setting failed

Error ID: 7310
The intensity maximum was not found during wavelength calibration.
Calibration 0 Failed:
Calibration 1 Failed:
Probable cause
1
2
3
4
5
6
7
Agilent InfinityLab LC Series VWD User Manual
Lamp is OFF.
Incorrect flow cell installation.
Flow cell contamination or air bubbles.
Intensity too low.
Current step value too far from maximum.
Misaligned/defective grating assembly.
Defective main board.
Zero-order calibration failed.
656 nm calibration failed.
Suggested actions
Switch on the lamp.
Ensure the flow cell is installed correctly.
Clean/replace flow cell windows or remove air
bubbles.
Replace lamp.
Repeat the calibration.
Please contact your Agilent service represen-
tative.
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.