Error Information
Detector Error Messages
No heater current
Error ID: 7453
The lamp heater current in the detector is missing. During lamp ignition,
the processor monitors the heater current. If the current does not rise
above the lower limit within 1 , the error message is generated.
Probable cause
1
2
3
4
5
Wavelength calibration setting failed
Error ID: 7310
The intensity maximum was not found during wavelength calibration.
Calibration 0 Failed:
Calibration 1 Failed:
Probable cause
1
2
3
106
Lamp disconnected.
Ignition started without the top foam in
place.
Defective main board.
Defective or non-Agilent lamp.
Defective power supply.
Lamp is OFF.
Incorrect flow cell installation.
Flow cell contamination or air bubbles.
Suggested actions
Ensure the lamp is connected.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Exchange the lamp.
Please contact your Agilent service
representative.
Zero-order calibration failed.
656 nm calibration failed.
Suggested actions
Switch on the lamp.
Ensure the flow cell is installed correctly.
Clean/replace flow cell windows or remove air
bubbles.