Error Information
Detector Error Messages

Wavelength Recalibration Lost

Error ID: 1037
The calibration information needed for your detector to operate correctly has
been lost.
During calibration of the detector the calibration values are stored in ROM. If no
data is available in the spectrometer ROM, the error message is generated.
Probable cause
1
2
Illegal Temperature Value from Sensor on Main
Board
Error ID: 1071
This temperature sensor (located on the detector main board) delivered a value
outside the allowed range. The parameter of this event equals the measured
temperature in 1/100 centigrade. As a result the temperature control is switched
off.
Probable cause
1
2
Agilent InfinityLab LC Series Diode Array Detectors User Manual
The detector is new.
The detector has been repaired.
Defective sensor or main board.
Detector is exposed to illegal ambient condi-
tions.
Suggested actions
Recalibrate the detector.
Please contact your Agilent service representa-
tive.
Suggested actions
Please contact your Agilent service representa-
tive.
Verify that the ambient conditions are within the
allowed range.
119