Wavelength Calibration Failed

This message may show up during a wavelength calibration.
If the expected deviation is larger than the specified wavelength accuracy, the
message
a Not Ready condition.
Probable cause
1
2
3
4

Wavelength Calibration Lost

After exchanging the monochromator assemblies, the calibration factors
should be reset to defaults values (a new FLM board comes with default
values). In this case
instrument stays in a Not Ready condition.
Probable cause
1
2
Agilent 1260 FLD User Manual
"Wavelength Calibration Failed"
Flash lamp not ignited or position not
correct.
Cell position not correct.
Solvent in the cell not clean or air bubble in
the cell.
monochromator assembly position not
correct (after replacement).
"Wavelength Calibration Lost"
Reset of monochromator settings after
exchange.
Replacement of FLM board.
Detector Error Messages
is displayed and the instrument stays in
Suggested actions
Check the flash lamp image and position. See
also
"No Peaks"
on page 142.
Check the cell position.
Flush the flow cell.
Reset monochromator settings and re-run the
wavelength calibration.
is displayed and the
Suggested actions
Perform a wavelength calibration.
Perform a wavelength calibration.