Error Information

Module Specific Error Messages

Module Specific Error Messages

These errors are detector-specific.

Internal temperature exceeded lower limit

Error ID: 10
Air temperature inside the instrument is ≤10 °C
Probable cause
1

Internal temperature exceeded upper limit

Error ID: 11
Air temperature inside the instrument is >40 °C
Probable cause
1

On-board Vapor sensor failed

Error ID: 12
The vapor sensor located on the main control board has failed.
Probable cause
1
2
Agilent InfinityLab ELSD User Manual
The environmental temperature is outside
the specified operating limits of the
instrument.
The environmental temperature is outside
the specified operating limits of the
instrument.
Vapor sensor not connected to the
mainboard.
Defective vapor sensor.
Suggested actions
Increase ambient temperature where
detector is located.
Move the detector to a warmer location.
Suggested actions
Decrease ambient temperature where
detector is located.
Move the detector to a cooler location.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.