Constancy Test
Images
Viewing the measurement re-
port
F.3-30
syngo CT 2007E
02/2008

Output of results

A defined number of results are stored. Older values are over-
written by new values.
The images recorded during measurement are stored as images
F . 3
of the "Quality Assurance Patient".
You can view the results of the daily and monthly quality test
with the Report function of the Local Service.
F . 3
Call up Options > Service > Local Service, delete the pass-
word entries, and click on OK.
Click on Reports in the Home Menu dialog window.
The Report Files dialog box is displayed.
F . 3
Quality Assurance
Emotion Operator Manual
F . 3
F . 3
F . 3
F . 3
F . 3