Instrument Location; Connect Power - Teledyne CombiFlash Rf User Manual

Table of Contents

Advertisement

+
CombiFlash® Rf
User Manual

4.2 Instrument Location

The CombiFlash Rf system requires approximately 1550
square centimeters (240 in
the CombiFlash Rf has at least 3 cm (1.25") of air space behind
it for ventilation. Additional space may be required for solvent
and waste containers.
The optional 4x expansion module is installed to the left of the
CombiFlash Rf system. If paired with the Rf
spectrometer, the 4x expansion module must be installed to
the right of the PurIon. The 4x module requires an additional
2
933 cm
has at least 3 cm (1.25") of air space behind it for ventilation.
Refer to Table 1-4 for environmental conditions and power
requirements.
Before making any connections to the CombiFlash Rf 4x
module, place it on the bench or in the fume hood where it will
be operated. Temporarily position the system so you can
access the back panels to complete the connections.

4.3 Connect Power

Ensure that the On/Standby switch below the CombiFlash Rf
system touch screen panel is in the Standby position. Then,
use the supplied IEC power cord to connect the CombiFlash Rf
4x module to mains power.
Mains power must meet the voltage, frequency, and amperage
requirements listed on the serial number label.
As long as the AC mains power cord is connected, power is
inside the unit. The mains power cord is the disconnect device.
Position the CombiFlash Rf system and Rf 4x module so that
the power cord can be unplugged, or use a power strip where
the plug can quickly be removed from the outlet in the event
of an emergency.
4-2
2
(145 in
) of level bench space. Ensure that the module
2
) of level bench space. Ensure that
WARNING
WARNING
+