3.5 Connect Waste Lines

Risk of fire or equipment damage. Failure to connect Waste
Port tubing may allow organic solvents to pool in unsafe areas,
possibly creating dangerous levels of flammable vapors.
Risk of fire ignited by electrostatic discharges. Never
substitute the black tubing on CombiFlash systems. The black
tubing (P/N 023-0503-06) is conductive. This tubing is required
to dissipate static electricity.
Elevated flammable vapor levels are possible. Ensure that the
waste container is adequately ventilated, preferably by placing it in
a fume hood.
Do not overfill the waste collection containers. Regularly monitor
and empty the bottles before they overfill.
The system has one waste outlet on the back panel: Diverter
Valve waste.
Accessory package (P/N 60-5239-012) includes a Waste Tube
Assembly (P/N 60-5234-187). To connect the Waste tube:
1. Connect the cut end of this assembly to the Diverter Valve
Waste port on the system:
a. locate the black nut and ferrule in the accessory
package. Slide the nut and ferrule onto the tubing as
shown in Figure 3-3.
3-6
WARNING
WARNING
CAUTION
CAUTION
Ferrule
Nut
Tubing