Table of Contents

Advertisement

+
CombiFlash® Rf
User Manual
The calibration routine places a series of targets on the touch
screen display. Touch the exact center of each target with a
stylus. The routine will move clockwise through 20 calibration
targets, after which the system will return to the normal
display.
During calibration routine, ensure that no part of your hand or
anything other than the stylus touches the screen.
Note
Note: If you select this command from a remote connection, the
system will start the calibration routine. Invoking this from a
remote connection may be useful if the system is miscalibrated
that this menu command cannot be selected from the
touchscreen.

6.2.5 Help

PeakTrak's Help menu has the following options:
6.2.5.1 Help
Choose this command to view the Help documentation.
6.2.5.2 Service
Qualified service personnel use this command to access the
system's service functions. These service functions are
password protected to restrict their use to only trained
personnel.
6.2.5.3 Software Update
Choose this command to open a window used to install patch
files to update your CombiFlash Rf system software.
6.2.5.4 Export Log Files
During operation, the system writes operating data to a log
file. Teledyne Isco service personnel can interpret this data to
optimize system performance or to troubleshoot difficulties.
Teledyne Isco recommends that you use the menu command
only when advised by a qualified service technician.
6.2.5.5 Last Reorder Code
This command should only be used by RediSep Rf Column
Plan customers. Displays the last reorder code generated by
the column plan. Use this to verify the reorder code was
6-10

Advertisement