DECLARATION OF CONFORMITY
Application of Council Directive:
Manufacturer's Address:
Equipment Type/Environment:
Trade Name/Model No:
Standards to which Conformity is Declared:
Standard
EN61000-4-2
Electrostatic Discharge
EN61000-4-3
Radiated RF Immunity
EN61000-4-4
Electrical Fast Transient
EN61000-4-5
Surge on AC Lines
EN61000-4-6
Conducted RF on AC lines
EN61000-4-11
Voltage Dips
Voltage Interruption
CISPR11/
RF Emissions
EN 55011
EN61000-3-2, 3-3
Harmonic and Flicker
I
The undersigned, hereby declares that the design of the equipment specified above confo rms to the above Directive(s) and Standards
as of September 5, 2008.
USA Representative
60-5232-505
2004/108/EC-The EMC Directive
2006/95/EC- The Low Voltage Directive
2002/96/EC – The WEEE Directive
Manufacturer's Name:
Teledyne Isco, Inc.
4700 Superior, Lincoln, Nebraska 68504 USA
P.O. Box 82531, Lincoln, NE 68501
Phone:
Facsimile: 1+ (402) 465-3799
Laboratory Equipment for Light Industrial/Commercial Environments
Combiflash RF4X
Year of Issue:
2008
EN 61010-1: 2
EN 61326:2006 - EMC Requirements for Electrical Equipment for
Description
William Foster
Vice President of Engineering
1+ (402) 464-0231
nd
Edition - Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
Measurement, Control, and Laboratory Use
Severity Applied
Level 2 - 4kV contact discharge
Level 3 - 8kV air discharge
80 Mhz to 2.7Ghz 80% AM at 1kHz
Level 2 - 3V/m
Level 2 - 1kV on AC lines
Level 2 - 2kV common mode,
Level 2 - 1KV differential mode
150 kHz to 80 MHz,
Level 1 - 1V rms, 80% modulated
1,10,25 cycles, 100,40,70%
250 cycles, 100% drop
Group 1, Class A Industrial, Scientific, and
Medical Equipment
Performance Criteria
A
A
A
A
A
A
C