Table 1-4 CombiFlash Rf + 4x System
Electrical Safety per EN 61010-1
Pollution Degree
Installation Category
Maximum Altitude
Note 1. All specifications are subject to change.
Note 2. For columns undetected by RFID, the max pressure is limited to 50 psi.
Table 1-5 CombiFlash EZ Prep Specifications
Dimensions
Weight
Power Options
Line Frequency
Ambient T emperature
Humidity (when
connected to power)
Flow Rate Range
Flow Rate Accuracy
b
Pressure Limit
Pressure Accuracy
Gradient Formation
Gradient Accuracy
Specifications
2
II
2000 meters
67 cm (26.5")  43 cm (17")  43 cm (17") (H  W  D)
37.4 kg (82.5 lbs.),
43.6 kg (96.1 lbs.) with ELSD detector
Input voltage range from 100 to 240 VAC, 50/60 Hz,
300VA maximum typical.
Line cord is the disconnect device.
50/60 Hz
20 to 40 °C (maximum temperature must be at least
15 °C below the boiling point of the solvent)
0 to 95% non-condensing
5 to 200 mL/min
±5% (5-200 mL/min)
241 bar (3500 psi)
5% of full scale
Binary gradient from solvent inlets 1-4. Solvent
selection can be automatically controlled by the
method.
±3% (20-200 mL/min)
1