CombiFlash® Rf
User Manual
Power Options
Line Frequency
Ambient Temperature
Humidity (when
connected to power)
Flow Rate Range
Flow Rate Accuracy
2
Pressure Limit
Pressure Accuracy
Gradient Formation
Peak Detection Modes
Flow Cell Path Length
UV Detection and
Monitor Wavelength
Wavelength Accuracy
Fraction Accuracy
1-14
Table 1-4 CombiFlash Rf + 4x System
Specifications
Input voltage range from 100 to 240 VAC, 50/60 Hz,
2.5 Amps system maximum, 0.5 amps module
maximum.
Line cord is the disconnect device.
50/60 Hz
20 to 40 °C (maximum temperature must be at least
15 °C below the boiling point of the solvent)
95% relative humidity maximum at 20 to 40 °C
1 to 200 mL/min
±5% (5-200 mL/min)
1379 kPa (200 psi)
5% of full scale
Binary gradient from A1, A2, and B1, B2 solvent inlets.
Solvent selection can be automatically controlled by
the method.
Slope or threshold
0.1 mm, ±25%
Selectable from 200 to 400 nm range. Optional 200 to
800 nm UV-Vis detector.
±5 nm
±[2mL  (flow rate  60)]
1
(Continued)