Teledyne CombiFlash Rf User Manual page 92

Table of Contents

Advertisement

Note
If the waste is to be collected in a common container, skip the
Column Waste and Waste Level Sense connections (steps 1
through 7) and complete step 8 to route the Valve Waste Out from
the Rf 4x module to the common container. The waste tubing
assembly connected to the back panel of the CombiFlash Rf
system will route all waste fluids to the common collection
container and sense the level.
To complete the waste line connections:
1. At the waste container for the CombiFlash Rf system
(Channel 1), remove the black tubing that delivers Diverter
Valve Waste fluid and the natural tubing that delivers
Waste Level Sense air. The cap and black tubing from the
Inject Valve Waste should remain attached to the waste
bottle.
2. At the CombiFlash Rf system back panel, disconnect the
black tubing from the Diverter Valve Waste port and the
natural tubing from the Waste Level Sense port.
Note
The Inject Valve Waste tubing must remain connected between
the Rf system back panel and its waste container.
Refer to Figures 4-5 and 4-6 for steps 3 and 4.
3. Locate the 20 inch (50 cm) long Column Waste Assembly
(P/N 60-5234-192) in accessory package 60-5239-008.
Connect this tubing between the CombiFlash Rf system
and the Rf 4x module:
a. Connect the cut end of the black tubing to the
CombiFlash Rf system Diverter Valve Waste port. Use a
black nut and ferrule from the accessory package.
Finger-tighten the nut to seat the ferrule in the
bulkhead fitting.
Section 4 Preparation: Optional 4x Module
4-7