DECLARATION OF CONFORMITY
Application of Council Directive:
Manufacturer's Name:
Manufacturer's Address:
Equipment Type/Environment:
Trade Name/Model No:
Standards to which Conformity is Declared:
Standard
EN61000-4-2:2008
Electrostatic Discharge
EN61000-4-3:2006
Radiated RF Immunity
/A1:2007 /A2:2010
EN61000-4-4:2004
Electrical Fast Transient (EFT) on
/A1:2010
Mains and I/O
EN61000-4-5:2005
Surge on AC Lines
EN61000-4-6:2008
Conducted RF Immunity on Mains and
I/O
EN61000-4-11:2004
Voltage Dips
CISPR11/
RF Emissions Radiated, below 1GHz
EN 55011:2009
and Conducted, AC Mains
/A1:2010
EN61000-3-2:2005
AC Harmonics, Flicker
/A1:2008 /A2:2009
EN61000-3-3:2008
We, the undersigned, hereby declare that the design of the equipment specified above conforms to the above Directive(s) and
Standards as of March 13. 2013.
USA Representative
_______________________________
Vikas V. Padhye Ph. D.
Vice President and Product General Manager
4700 Superior Street
Lincoln, Nebraska 68504
Phone: 402-464-0231
Fax:
402-464-0318
60-5232-509
Rev. B
2004/108/EC -The EMC Directive
2002/96/EC
The WEEE Directive
2006/95/EC
The Low Voltage Directive
Teledyne Isco
4700 Superior, Lincoln, Nebraska 68504 USA
Mailing Address: P.O. Box 82531, Lincoln, NE 68501
Phone: +1 (402) 464-0231
FAX:
+1 (402) 465-3799
Laboratory Equipment for Light Industrial/Commercial Environments
Rf 150
Year of Issue:
2013
EN 61326:2006 EMC Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use
EN61010-1:2010 Safety Requirements for Electrical Equipment for Measurement, Control,
and Laboratory Use.
Description
Level 2 - 4kV contact discharge
Level 3 - 8kV air discharge
80 mHz to 2.7gHz 80% AM at 1 kHz
Level 2 - 3V/m
Level 2 - 1kV on AC lines
Level 2 - 1kV Line Common Mode
Level 2 - 0.5kV Differential Mode
150 kHz to 80 mHz
Level 1
0% during1 cycle and full cycle
70% at 25 cycles
Group 1, Class A Industrial, Scientific, and
Medical Equipment
Severity Applied
1V rms, 80% Modulated
Performance Criteria
A
A
A
A
A
A
PASS