8.1 System Standby and Shut Down

During extended periods of inactivity, you can place the
system in Standby to conserve power. To do so, log off the
system (File > Log out) and place the
Standby.
When in the Standby state, normal system operation is no
longer available from the touch screen or remotely. However,
some internal components are still powered.
As long as the AC mains power cord is connected, power is
inside the unit. The mains power cord is the disconnect device.
Position the CombiFlash Rf system so that the power cord can
be unplugged, or use a power strip where the plug can quickly
be removed from the outlet in the event of an emergency.
When you first place the system in Standby, internal
components continue to operate for almost one minute while
performing file maintenance and preparing the system for
possible power removal.
Removing the AC mains power cord before the file maintenance is
complete might corrupt files on the internal hard drive. These
corrupted files can cause abnormal operation or a complete
system failure that requires service. Unless power must be
removed due to an emergency, always wait at least one minute
after placing the system in Standby before removing the AC mains
power cord.