Section 2 Preparation: Rf + , Rf + Lumen, Rf+ PurIon, and EZ Prep Systems

2.6 EZ Prep Plumbing Connections

The EZ Prep comes with the solvent inlet lines and waste lines
pre-assembled to the instrument. To install the EZ Prep:
1. Remove the shipping clamps that retain the plumbing lines
(Figure 2-4).
Figure 2-4 Remove the two shipping clamps indicated above
2. Remove the 2 screws and tubing clamps identified in figure
2-4. This releases the tubing lines for placement in the
solvent supply and waste reservoirs.
Note
Due to the large size of the inlet filter and inlet line tubing
connections, the tubing assemblies have been pre-assembled with
a bottle cap the is typical for the region (GL 38 in the USA and GL
45 in the rest of the world). If these caps aren't suitable for your
containers, you will have to make modifications as needed.
3. Place the solvent lines in the solvent supply containers.
The solvent inlet filters will need to be placed into the