After you have prepared the pre-filled or empty cartridge,
place the solid sample cartridge on the system:
1. Attach the desired adjustable cartridge cap:
• 60-5237-047, fits 2.5 and 5 gram solid load cartridges.
• 60-5237-048, fits 12 and 20 gram solid load
cartridges.
• 60-5237-044, fits 32 and 65 gram solid load
cartridges.
2. Press the lever on the side of the cap and fully extend the
plunger (Figure 5-1).
1
Figure 5-1 Press the lever and extend the plunger
3. Slide the plunger into the cartridge until it reaches the top
frit (Figure 5-2).
4. Press the lever and push the cartridge into the cap. Align
the cartridge so that it fits fully into the recess in the cap
(Figure 5-3).
5. Load the solid sample cartridge with cap on the sample
injection port.
2
Section 5 Operation
5-3