Connect Waste Lines - Teledyne CombiFlash Rf User Manual

Table of Contents

Advertisement

4x module
Figure 4-3 Coupler for connecting 4x tubing
4. Connect the other end of the shorter tubing assembly to the
tee fitting on the back of the PurIon.
5. Connect the longer tubing assembly to the "D" port on the
Rf and to the other side of the tee fitting on the back of the
PurIon.

4.5 Connect Waste Lines

Risk of fire or equipment damage. Failure to connect Waste
Port tubing may allow organic solvents to pool in unsafe areas,
possibly creating dangerous levels of flammable vapors.
Risk of fire ignited by electrostatic discharges. Never
substitute the black tubing on CombiFlash systems. The black
tubing (P/N 023-0503-06) is conductive. This tubing is required
to dissipate static electricity.
Elevated flammable vapor levels are possible. Ensure that the
waste container is adequately ventilated, preferably by placing it in
a fume hood.
The Rf 4x module has six liquid waste ports on the back panel
(Figure 4-4):
Section 4 Preparation: Optional 4x Module
WARNING
WARNING
CAUTION
PurIon
4-5