SRCPOFS Source Power Offset

SRCPOFS Source Power Offset
Syntax
Figure 7-251
SRCPOFS Syntax
Description
The SRCPOFS command offsets the displayed power of the built-in
tracking generator. This function can be used to take into account
system losses (for example, cable loss) or gains (for example,
preamplifier gain) reflecting the actual power delivered to the device
under test. SRCPOFS is available only with an 8560E/EC Option 002.
Parameters
number
UP or DN
Preset State
0.0 dB
604
real from −100 dB to +100 dB.
increments in steps equal to the value set by SRCPSTP.