CARRON Carrier On Power

Syntax
Figure 7-70
CARRON Syntax
Description
The CARRON command measures the average power of the carrier
during that portion of the time when it is on (when it is within 20 dB of
its peak level). True mean carrier power is calculated by measuring the
time waveform of the RF envelope, converting the trace data from dB to
power units, and then averaging the power trace data.
The MEANPWR command makes a similar measurement but it allows
you to define the on power range.
Query Response
Figure 7-71
CARRON Query Response
Example
10
20
30
40
Chapter 7
REAL Carron_pwr
OUTPUT 718;"CARRON TRB,?;"
ENTER 718;Carron_pwr
END
Language Reference
CARRON Carrier On Power
433