AUTOCPL Auto Coupled

AUTOCPL Auto Coupled
Syntax
Figure 7-64
AUTOCPL Syntax
Description
The AUTOCPL command sets video bandwidth, resolution bandwidth,
input attenuation, sweep time, and center frequency step-size to
coupled mode. These functions can be recoupled individually or all at
once. The spectrum analyzer chooses appropriate values for these
functions. The video bandwidth and resolution bandwidth are set
according to the coupled ratios stored with VBW/RBW RATIO or
RBW/SPAN RATIO, or the ratios selected using the VBR or RBR
commands. If no ratios are chosen, default ratios (1.0 and 0.011,
respectively) are used instead.