Language Reference

AMPCOR Amplitude Correction

AMPCOR Amplitude Correction
Syntax
Figure 7-49
AMPCOR Syntax
Description
Use AMPCOR to turn the amplitude correction function on and off. The
ampcor function is used to compensate for frequency-dependent
amplitude variations. When ampcor is on, the current correction values
are added to all measurement results. Turning ampcor off does not
erase the current frequency-amplitude correction factors. Performing
an instrument preset (IP) or turning off the spectrum analyzer will turn
off ampcor.
Preset State
Off
Query Response
Figure 7-50
AMPCOR Query Response
Example
10
416
OUTPUT 718;"AMPCOR ON;"