Figure 2-78 Resolution Bandwidth Filter Charge-Up Effects - Agilent Technologies 8560E User Manual

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

Figure 2-78
Resolution Bandwidth Filter Charge-Up Effects
Because the resolution-bandwidth filters are band-limited devices,
they require a finite amount of time to react to changing conditions.
Specifically, the filters take time to charge fully after the analyzer is
exposed to a pulsed signal. Although the charge-up is asymptotic,
you can consider it to be complete after the length of time defined by
two divided by resolution bandwidth, where the resolution
bandwidth is the −3 dB resolution bandwidth being used.
Because setup time should be greater than filter charge times, be
sure that:
or
where, in this example, RBW is greater than 2/1.5 ms; that is, RBW
is greater than 1333 Hz. The resolution bandwidth should be set to
the next larger value, 3 kHz.
Chapter 2
Example 11: Time-Gated Measurement
2
>
SUT
------------ -
RBW
2
>
RBW
-----------
SUT
Making Measurements
147