Language Reference
MKFCR Frequency Counter Resolution
Example
10
INPUT "ENTER IN THE DESIRED CENTER FREQUENCY, IN MHZ",F
req
20
INPUT "ENTER IN THE DESIRED FREQUENCY SPAN, IN MHZ",Spa
n
30
OUTPUT 718;"IP;CF ";Freq;"MHZ;"
40
OUTPUT 718;"SP ";Span;"MHZ;"
50
INPUT "ENTER DESIRED FREQUENCY-COUNTER RESOLUTION, IN
HZ",Resolution
60
OUTPUT 718;"MKFCR ";Resolution;"HZ;",
70
OUTPUT 718;"MKN EP;"
80
PRINT "PLACE THE MARKER ON THE DESIRED SIGNAL."
90
PRINT "PRESS HOLD ON THE ANALYZER, THEN PRESS CONTINUE.
"
100
PAUSE
110
OUTPUT 718;"MKFC ON;"
120
OUTPUT 718;"MKF?;"
130
ENTER 718;Freq_count
140
PRINT "FREQUENCY IS",Freq_count,"HZ"
150
END