Agilent Technologies 8560E User Manual page 699

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

AMPL 1M Unable to adjust amplitude of 1 MHz RES BW.
ERR 507
AMPL 30K Insufficient gain during LC BW Cal of 30 kHz
ERR 508
RES BW.
AMPL .1M Insufficient gain during LC BW Cal of 100 kHz
ERR 509
RES BW.
AMPL .3M Insufficient gain during LC BW Cal of 300 kHz
ERR 510
RES BW.
AMPL 1M Insufficient gain during LC BW Cal of 1 MHz
ERR 511
RES BW.
RBW <300 Insufficient gain during crystal BW Cal of less
ERR 512
than 300 Hz RES BW.
RBW 300 Insufficient gain during crystal BW Cal of 300
ERR 513
Hz RES BW.
RBW 1K Insufficient gain during crystal BW Cal of 1 kHz
ERR 514
RES BW.
RBW 3K Insufficient gain during crystal BW Cal of 3 kHz
ERR 515
RES BW.
RBW 10K Insufficient gain during crystal BW Cal of 10
ERR 516
kHz RES BW.
RBW 100 Unable to adjust 100 Hz RES BW.
ERR 517
RBW 300 Unable to adjust 300 Hz RES BW.
ERR 518
RBW 1K Unable to adjust 1 kHz RES BW.
ERR 519
RBW 3K Unable to adjust 3 kHz RES BW.
ERR 520
RBW 10K Unable to adjust 10 kHz RES BW.
ERR 521
RBW 10K Unable to adjust 10 kHz RES BW symmetry of
ERR 522
pole 1.
RBW 10K Unable to adjust 10 kHz RES BW symmetry of
ERR 523
pole 2.
RBW 10K Unable to adjust 10 kHz RES BW symmetry of
ERR 524
pole 3.
RBW 10K Unable to adjust 10 kHz RES BW symmetry of
ERR 525
pole 4.
RBW <300 Unable to adjust less than 300 Hz RES BWs.
ERR 526
Timeout during data sampling.
RBW <300 Step gain correction failed for less than 300 Hz
ERR 527
RES BW.
Chapter 9
If You Have a Problem
Error Messages
699