Agilent Technologies 8560E User Manual page 692

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

If You Have a Problem
Error Messages
Error codes 200 through 299 relate to ADC hardware/firmware
failures. Instrument service is required.
SYSTEM ADC Driver/ADC hardware/firmware interaction;
ERR 200
check for other errors.
SYSTEM ADC Controller/ADC hardware/firmware
ERR 201
interaction; check for other errors.
FADC CAL Linear offset search failed.
ERR 202
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
FADC CAL Log offset search failed.
ERR 203
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
FADC CAL Log 2× expand offset search failed.
ERR 204
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
FADC CAL Linear slope calibration is out of range.
ERR 205
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
FADC CAL Log slope calibration is out of range.
ERR 206
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
FADC CAL Log 2× expand slope calibration is out of range.
ERR 207
This error applies applies only to EC-series analyzers
and E-series analyzers with fast ADC (Option 007)
OUTOF RG ADC input is outside of ADC range.
ERR 250
NO IRQ Microprocessor not receiving interrupt from ADC.
ERR 251