7 mm verification kit

Advertisement

Quick Links

User's and Service Guide
Agilent Technologies 85051B
7 mm Verification Kit
This manual applies directly to 85051B Verification Kits that have serial number prefix 2815A.
The verification devices in this kit are individually serialized.
Manufacturing Part Number: 85051-90031
Printed in USA
Print Date: August 2012
Supersedes: July 2007
© Copyright 1995, 2002, 2007, 2012 Agilent Technologies, Inc. All rights reserved.

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 85051B

  • Page 1 User’s and Service Guide Agilent Technologies 85051B 7 mm Verification Kit This manual applies directly to 85051B Verification Kits that have serial number prefix 2815A. The verification devices in this kit are individually serialized. Manufacturing Part Number: 85051-90031 Printed in USA...
  • Page 2 Certification Agilent Technologies, Inc. certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies, Inc. further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members.
  • Page 3 To print copies of documentation from the Web, download the PDF file from the Agilent web site: • Go to www.agilent.com. • Enter the document’s part number (located on the title page) in the Search box. • Click Search. • Open the PDF and print the document. 85051B...
  • Page 4 85051B...
  • Page 5: Table Of Contents

    Where to Look for More Information ............... . . 5-4 85051B...
  • Page 6 Introduction....................6-2 85051B...
  • Page 7: General Information

    General Information 85051B...
  • Page 8: Verification Kit Overview

    5-5. Calibration Definitions The 85051B verification kit is intended to be used with the 85050B/C/D 7 mm calibration kits. Prior to performing a calibration with your PNA, the calibration kit must be selected and the calibration definitions for the devices in the kit installed in the PNA. Refer to your PNA Help system for instructions on selecting the calibration kit and performing a calibration.
  • Page 9: Equipment Required But Not Supplied

    Check for damage. The foam-lined storage case provides protection during shipping. Verify that this case and its contents are not damaged. If the case or any device appears damaged, or if the shipment is incomplete, contact Agilent Technologies. “Contacting Agilent” on page 5-5.
  • Page 10: Recording The Device Serial Numbers

    Record these serial numbers in Table 1-1. Recording the serial numbers will prevent confusing the devices in this kit with similar devices from other kits. Table 1-1 Serial Number Record for the 85051B Device Serial Number Verification kit...
  • Page 11: Preventive Maintenance

    Improper connections, resulting from pin depth values being out of specification (see Table 2-2 on page 2-4), or from bad connection techniques, can also damage these devices. 85051B...
  • Page 12 General Information Clarifying the Terminology of a Connector Interface 85051B...
  • Page 13: Specifications

    Specifications 85051B...
  • Page 14: Environmental Requirements

    Due to the small dimensions of the devices, electrical characteristics will change with temperature. Therefore, the operating temperature is a critical factor in their performance, and must be stable before use. IMPORTANT Avoid unnecessary handling the devices during use because your fingers are a heat source. 85051B...
  • Page 15: Mechanical Characteristics

    They are, however, important supplemental characteristics related to electrical performance. Agilent Technologies verifies the mechanical characteristics of the devices in this kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any improper pin depth when the kit leaves the factory.
  • Page 16: Airline Mechanical Characteristics

    The center and outer conductors of the airlines in this kit have been mechanically measured and matched. Do not use the center or outer conductors provided in this kit with a center or outer conductor from any other airline. Damage to the airline or attaching connector may result. 85051B...
  • Page 17 Dimension millimeters inches 3.04 ±0.0025 0.1197 ±0.00010 L +0.0025/0.010 +0.00010/0.0004 Figure 2-4 25 Mismatch Airline Center Conductor Dimension millimeters inches 3.04 ±0.008 0.1197 ±0.0003 4.613 ±0.005 0.1816 ±0.0002 74.93 ±0.019 2.950 ±0.0007 12.514 ±0.050 0.4927 ±0.0020 L +0.0025/0.0100 +0.0001/0.0004 85051B...
  • Page 18: Electrical Specifications

    At the factory, each verification device is electrically characterized on a PNA measurement system. These factory measurements are traceable to the National Institute of Standards and Technology (NIST) through mechanical and electrical paths (for more information on traceability, contact Agilent Technologies. Refer to “Contacting Agilent” on page 5-5.
  • Page 19: Use, Maintenance, And Care Of The Devices

    Use, Maintenance, and Care of the Devices 85051B...
  • Page 20: Electrostatic Discharge

    4. Remove the short from the cable. Figure 3-1 shows a typical ESD protection setup using a grounded mat an wrist strap. Refer to Table 6-2 on page 6-3 for information on ordering supplies for ESD protection. Figure 3-1 ESD Protection Setup 85051B...
  • Page 21: Visual Inspection

    Other small defects and cosmetic imperfections are also normal. None of these affect electrical or mechanical performance. If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. 85051B...
  • Page 22: Cleaning Connectors

    Refer to Table 6-2 on page 6-3 for a part number for cleaning swabs. a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab. b. Clean the connector threads. 85051B...
  • Page 23 Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it. 4. Inspect Inspect the connector to make sure that no particles or residue remain. Refer to “Visual Inspection” on page 3-3. 85051B...
  • Page 24: Gaging Connectors

    Gaging Connectors Gaging Connectors The gages available from Agilent Technologies are intended for preventive maintenance and troubleshooting purposes only. They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on DUTs, test accessories, and the calibration kit devices.Do not use the gages for precise pin depth measurements.
  • Page 25: Reading The Connector Gage

    If the center conductor protrudes, the indicator will move clockwise to indicate the amount of protrusion, which is read as a positive value. Figure 3-2 Reading the Connector Gage 85051B...
  • Page 26: Gaging Procedure

    For maximum accuracy, measure the connector a minimum of three times and take an average of the readings. After each measurement, rotate the gage a quarter-turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis. 85051B...
  • Page 27 Use, Maintenance, and Care of the Devices Gaging Connectors e. Compare the average reading with the pin depth specifications listed in Table 2-2 on page 2-4. f. Remove the device from the gage. Figure 3-3 Gaging 7 mm Connectors 85051B...
  • Page 28: Gaging The Airline

    “A” end of the outer conductor. Do not let the center conductor scrape the edge of the outer conductor or damage may result. 8. Attach the gage to the other end of the airline. Refer to Figure 3-5. 3-10 85051B...
  • Page 29 Replace the other plastic cap on the outer conductor and store the center and outer conductors in the foam lined storage case. 85051B 3-11...
  • Page 30: Making Connections

    A torque wrench is included in the calibration kit. Refer to Table 6-1 on page 6-2 for part number and ordering information. Table 3-1 Torque Wrench Information Connector Type Torque Setting Torque Tolerance 13.6 N-cm (1.2 in-lb) 7 mm 136 N-cm (12 in-lb) 3-12 85051B...
  • Page 31 3. Apply downward force perpendicular to the wrench handle. This applies torque to the connection through the wrench. Do not hold the wrench so tightly that you push the handle straight down along its length rather than pivoting it, otherwise you apply an unknown amount of torque. 85051B 3-13...
  • Page 32: Connecting The Airline

    Detachable spring-loaded tips are supplied with the center conductors of both the airline and mismatch airline in your 85051B kit. Do not use these tips when connecting the airlines to the TRL adapters supplied in the 85050C 7 mm Precision Calibration Kit.
  • Page 33 Airline NOTE To avoid damaging the airline center conductor, always keep it in line with the connecting center pin. 7. Insert the tip of the center conductor (emerging from the “A” end) into the cable center conductor. See 85051B 3-15...
  • Page 34 10. Mate the outer conductors of the port 2 cable connector and airline finger tight. 11. Torque the cable connectors as shown in Figure 3-11 on page 3-17. To keep the airline from turning, either hold its plastic insulation or use an open end wrench while you torque the connection. 3-16 85051B...
  • Page 35: How To Separate A Connection

    If disconnecting an airline and the airline center conductor does not disengage from the device center conductor, gently pull the center conductors apart and then push the airline center conductor back inside the outer conductor of the airline. 85051B 3-17...
  • Page 36: Handling And Storage

    • Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface. 3-18 85051B...
  • Page 37: Performance Verification

    Performance Verification 85051B...
  • Page 38: Introduction

    Introduction Introduction The performance of your verification kit can only be verified by returning the kit to Agilent Technologies for recertification. The equipment required to verify the specifications of the devices in the kit has been specially manufactured and is not commercially available.
  • Page 39: Recertification

    A list of NIST traceable numbers may be purchased upon request to be included in the calibration report. Agilent Technologies offers a Standard calibration for the recertification of the kit. For more information, contact Agilent Technologies. See “Contacting Agilent” on page 5-5.
  • Page 40 Performance Verification Recertification 85051B...
  • Page 41: Troubleshooting

    Troubleshooting 85051B...
  • Page 42: Troubleshooting Process

    Troubleshooting Troubleshooting Process Troubleshooting Process If your PNA does not pass performance verification, follow the steps in Figure 5-1 to determine the cause of the failure and the correct action to take to correct the failure. 85051B...
  • Page 43 Troubleshooting Troubleshooting Process Figure 5-1 Troubleshooting Flowchart 85051B...
  • Page 44: Compatible Network Analyzers

    5-5. Returning a Kit or Device to Agilent If your kit or device requires service, contact Agilent Technologies for information on where to send it. See “Contacting Agilent” on page 5-5. Include a service tag (located near the end of this manual) on which you provide the following information: •...
  • Page 45 If you do not have access to the Internet, please contact your Agilent field engineer. NOTE In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number. With this information, the Agilent representative can determine whether your product is still within its warranty period. 85051B...
  • Page 46 Troubleshooting Contacting Agilent 85051B...
  • Page 47 Replaceable Parts 85051B...
  • Page 48 Introduction Introduction Table 6-1 lists the replacement part numbers for items included in the 85051B verification kit. Table 6-2 lists the replacement part numbers for items not included in the verification kit that are either required or recommended for successful operation of the kit.
  • Page 49 Cleaning swabs a. Included in the 85050B and 85050C 7 mm calibration kits. Included in the 85050B, 85050C, and 85050D 7 mm calibration kits. c. Agilent can no longer safely ship isopropyl alcohol, so customers should purchase it locally. 85051B...
  • Page 50 Replaceable Parts Introduction Figure 6-1 Replaceable Parts 85051B...
  • Page 51 See calibration definitions constants, See calibration definitions contacting Agilent Technologies definitions cord entering grounding permanently stored part number kits intended to be used report temperature...
  • Page 52 Agilent verifies heel strap verifying precautions protection protection setup supplies maintenance part numbers agreements wrist strap of devices preventive making connections 3-12 ESD protection 3-12 flowchart precautions 3-12 troubleshooting manual frequency range part number conductive gage part number connector Index-2 85051B...