User’s and Service Guide Agilent Technologies 85051B 7 mm Verification Kit This manual applies directly to 85051B Verification Kits that have serial number prefix 2815A. The verification devices in this kit are individually serialized. Manufacturing Part Number: 85051-90031 Printed in USA...
Page 2
Certification Agilent Technologies, Inc. certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies, Inc. further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members.
Page 3
To print copies of documentation from the Web, download the PDF file from the Agilent web site: • Go to www.agilent.com. • Enter the document’s part number (located on the title page) in the Search box. • Click Search. • Open the PDF and print the document. 85051B...
5-5. Calibration Definitions The 85051B verification kit is intended to be used with the 85050B/C/D 7 mm calibration kits. Prior to performing a calibration with your PNA, the calibration kit must be selected and the calibration definitions for the devices in the kit installed in the PNA. Refer to your PNA Help system for instructions on selecting the calibration kit and performing a calibration.
Check for damage. The foam-lined storage case provides protection during shipping. Verify that this case and its contents are not damaged. If the case or any device appears damaged, or if the shipment is incomplete, contact Agilent Technologies. “Contacting Agilent” on page 5-5.
Record these serial numbers in Table 1-1. Recording the serial numbers will prevent confusing the devices in this kit with similar devices from other kits. Table 1-1 Serial Number Record for the 85051B Device Serial Number Verification kit...
Improper connections, resulting from pin depth values being out of specification (see Table 2-2 on page 2-4), or from bad connection techniques, can also damage these devices. 85051B...
Page 12
General Information Clarifying the Terminology of a Connector Interface 85051B...
Due to the small dimensions of the devices, electrical characteristics will change with temperature. Therefore, the operating temperature is a critical factor in their performance, and must be stable before use. IMPORTANT Avoid unnecessary handling the devices during use because your fingers are a heat source. 85051B...
They are, however, important supplemental characteristics related to electrical performance. Agilent Technologies verifies the mechanical characteristics of the devices in this kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any improper pin depth when the kit leaves the factory.
The center and outer conductors of the airlines in this kit have been mechanically measured and matched. Do not use the center or outer conductors provided in this kit with a center or outer conductor from any other airline. Damage to the airline or attaching connector may result. 85051B...
At the factory, each verification device is electrically characterized on a PNA measurement system. These factory measurements are traceable to the National Institute of Standards and Technology (NIST) through mechanical and electrical paths (for more information on traceability, contact Agilent Technologies. Refer to “Contacting Agilent” on page 5-5.
4. Remove the short from the cable. Figure 3-1 shows a typical ESD protection setup using a grounded mat an wrist strap. Refer to Table 6-2 on page 6-3 for information on ordering supplies for ESD protection. Figure 3-1 ESD Protection Setup 85051B...
Other small defects and cosmetic imperfections are also normal. None of these affect electrical or mechanical performance. If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. 85051B...
Refer to Table 6-2 on page 6-3 for a part number for cleaning swabs. a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab. b. Clean the connector threads. 85051B...
Page 23
Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it. 4. Inspect Inspect the connector to make sure that no particles or residue remain. Refer to “Visual Inspection” on page 3-3. 85051B...
Gaging Connectors Gaging Connectors The gages available from Agilent Technologies are intended for preventive maintenance and troubleshooting purposes only. They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on DUTs, test accessories, and the calibration kit devices.Do not use the gages for precise pin depth measurements.
If the center conductor protrudes, the indicator will move clockwise to indicate the amount of protrusion, which is read as a positive value. Figure 3-2 Reading the Connector Gage 85051B...
For maximum accuracy, measure the connector a minimum of three times and take an average of the readings. After each measurement, rotate the gage a quarter-turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis. 85051B...
Page 27
Use, Maintenance, and Care of the Devices Gaging Connectors e. Compare the average reading with the pin depth specifications listed in Table 2-2 on page 2-4. f. Remove the device from the gage. Figure 3-3 Gaging 7 mm Connectors 85051B...
“A” end of the outer conductor. Do not let the center conductor scrape the edge of the outer conductor or damage may result. 8. Attach the gage to the other end of the airline. Refer to Figure 3-5. 3-10 85051B...
Page 29
Replace the other plastic cap on the outer conductor and store the center and outer conductors in the foam lined storage case. 85051B 3-11...
A torque wrench is included in the calibration kit. Refer to Table 6-1 on page 6-2 for part number and ordering information. Table 3-1 Torque Wrench Information Connector Type Torque Setting Torque Tolerance 13.6 N-cm (1.2 in-lb) 7 mm 136 N-cm (12 in-lb) 3-12 85051B...
Page 31
3. Apply downward force perpendicular to the wrench handle. This applies torque to the connection through the wrench. Do not hold the wrench so tightly that you push the handle straight down along its length rather than pivoting it, otherwise you apply an unknown amount of torque. 85051B 3-13...
Detachable spring-loaded tips are supplied with the center conductors of both the airline and mismatch airline in your 85051B kit. Do not use these tips when connecting the airlines to the TRL adapters supplied in the 85050C 7 mm Precision Calibration Kit.
Page 33
Airline NOTE To avoid damaging the airline center conductor, always keep it in line with the connecting center pin. 7. Insert the tip of the center conductor (emerging from the “A” end) into the cable center conductor. See 85051B 3-15...
Page 34
10. Mate the outer conductors of the port 2 cable connector and airline finger tight. 11. Torque the cable connectors as shown in Figure 3-11 on page 3-17. To keep the airline from turning, either hold its plastic insulation or use an open end wrench while you torque the connection. 3-16 85051B...
If disconnecting an airline and the airline center conductor does not disengage from the device center conductor, gently pull the center conductors apart and then push the airline center conductor back inside the outer conductor of the airline. 85051B 3-17...
• Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface. 3-18 85051B...
Introduction Introduction The performance of your verification kit can only be verified by returning the kit to Agilent Technologies for recertification. The equipment required to verify the specifications of the devices in the kit has been specially manufactured and is not commercially available.
A list of NIST traceable numbers may be purchased upon request to be included in the calibration report. Agilent Technologies offers a Standard calibration for the recertification of the kit. For more information, contact Agilent Technologies. See “Contacting Agilent” on page 5-5.
Troubleshooting Troubleshooting Process Troubleshooting Process If your PNA does not pass performance verification, follow the steps in Figure 5-1 to determine the cause of the failure and the correct action to take to correct the failure. 85051B...
Page 43
Troubleshooting Troubleshooting Process Figure 5-1 Troubleshooting Flowchart 85051B...
5-5. Returning a Kit or Device to Agilent If your kit or device requires service, contact Agilent Technologies for information on where to send it. See “Contacting Agilent” on page 5-5. Include a service tag (located near the end of this manual) on which you provide the following information: •...
Page 45
If you do not have access to the Internet, please contact your Agilent field engineer. NOTE In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number. With this information, the Agilent representative can determine whether your product is still within its warranty period. 85051B...
Page 48
Introduction Introduction Table 6-1 lists the replacement part numbers for items included in the 85051B verification kit. Table 6-2 lists the replacement part numbers for items not included in the verification kit that are either required or recommended for successful operation of the kit.
Page 49
Cleaning swabs a. Included in the 85050B and 85050C 7 mm calibration kits. Included in the 85050B, 85050C, and 85050D 7 mm calibration kits. c. Agilent can no longer safely ship isopropyl alcohol, so customers should purchase it locally. 85051B...
Page 50
Replaceable Parts Introduction Figure 6-1 Replaceable Parts 85051B...
Page 51
See calibration definitions constants, See calibration definitions contacting Agilent Technologies definitions cord entering grounding permanently stored part number kits intended to be used report temperature...
Page 52
Agilent verifies heel strap verifying precautions protection protection setup supplies maintenance part numbers agreements wrist strap of devices preventive making connections 3-12 ESD protection 3-12 flowchart precautions 3-12 troubleshooting manual frequency range part number conductive gage part number connector Index-2 85051B...
Page 54
3-13 3-17 part number proper positioning of 3-13 torque 3-12 3-13 part number precautions for use of 3-13 proper use of 3-13 wrist strap for ESD protection part number zeroing connector gage Index-4 85051B...