Mxmh Maximum Hold; Figure 7-191 Mxmh Syntax - Agilent Technologies 8560E User Manual

Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

MXMH Maximum Hold

Syntax
Figure 7-191
MXMH Syntax
Description
The MXMH command updates the chosen trace with the maximum
signal level detected at each trace-data point from subsequent sweeps.
This function employs the positive peak detector (refer to the DET
command). The detector mode can be changed, if desired, after
maximum hold is initialized.
Example
10
20
Chapter 7
OUTPUT 718;"BLANK TRA;CLRW TRB;MXMH TRB;"
END