Agilent technologies 8560 e-series and ec-series spectrum analyzers
Table of Contents

Advertisement

Programming
Math Functions
E
2
XAMPLE
10
! PUT TRACES ON SCREEN
20
INTEGER Atrace(1:601)
30
FOR I=1 TO 601
40
Atrace(I)=300
50
NEXT I
60
OUTPUT 718;"IP;LG 10DB;SNGLS;TS;"
70
OUTPUT 718 USING "#,K,W,601(W) ,K";"TDF
A;TRA#A",1202,Atrace(*),";"
80
OUTPUT 718 USING "#,K,W,601(W) ,K";"TDF
A;TRB#A",1202,Atrace(*),";"
90
OUTPUT 718;"AMB ON;"
100
PRINT "PRESS CONTINUE"
110
PAUSE
120
OUTPUT 718;"LN;SNGLS;TS;"
130
OUTPUT 718 USING "#,K,W,601(W) ,K";"TDF
A;TRA#A",1202,Atrace(*),";"
140
OUTPUT 718 USING "#,K,W,601(W) ,K";"TDF
A;TRB#A",1202,Atrace(*),";"
150
OUTPUT 718;"AMB ON;"
160
PAUSE
170
END
Trace Data Limits
The displayed amplitude of each trace element falls in one of 601
vertical data points. See Figure 5-10. There are an additional 10 points
of overrange. The spectrum analyzer clips results that exceed these
limits. The overrange is equal to one-sixth of a division above the
reference level. Also, the same clipping algorithm is applied to
correction data in a trace (for example, correction data that you enter
into trace B). For example, if the reference level is 0 dBm, the scale is
equal to 10 dB per division; therefore, the correction values must be
within the range of +1.66 dBm to −100.00 dBm (one-sixth of 10 dB is
equal to 1.66 dB).
322