Eliminating Error Messages
It might be possible to eliminate some error messages by running the
LO and IF realignment procedure below, or by running the procedures
described in "Hardware Problems". If an error message remains
displayed, the following actions are suggested:
Error
Action
100 to 199
Programming error detected. Refer to chapters 3 and 7
in the user's guide for information on programming the
spectrum analyzer.
200 to 799
The instrument is in need of service. Refer to this
chapter for information about calling Agilent
Technologies.
800 to 899
Option module failures detected. Refer to the
appropriate option-module for an explanation of these
errors.
900 to 999
User generated errors. Information entered incorrectly
or an attempt to use the spectrum analyzer
inappropriately.
Realigning the LO and IF
The LO and IF realignment procedure may eliminate error conditions.
1. Press
, then
SAVE
2. Store the current state in a convenient STATE register.
3. Press
,
PRESET
CAL
finish.
4. Press
, then
RECALL
5. Recall the previously stored STATE.
Chapter 9
.
SAVE STATE
, then
REALIGN LO &IF
.
RECALL STATE
If You Have a Problem